Skip to content

Assembly overview

Nikita Vladimirov edited this page Sep 6, 2022 · 31 revisions

Assembly animation

The animation above shows you 4 major modules: left excitation arm, detection arm, sample stages, and right excitation arm, in recommended assembling sequence. This is only an overview - see the detailed instructions for each module for the latest components and their arrangement.

Left excitation arm

Optical design

Left excitation arm 2D

Laser light from the fiber output is collimated by L1 and then redirected into the electrically tunable lens (ETL). A 1:1 relay (lenses L2, L3) reimages the ETL onto a galvo mirror (M3), which is placed at the back pupil plane of the excitation objective L4 (modified f=50 mm Nikon objective). Scanning the Gaussian beam along one axis generates the light-sheet illumination. Modulation of the ETL optical power displaces the light-sheet axially (left to right), and it is synchronized with the sCMOS camera rolling shutter (ASLM mode). The fold mirrors M1, M2 make the system more compact and allow angular alignment of the laser beam.

Mechanical design

Left excitation arm 3D Excitation arms 3D rendering of the excitation arms. Parts that need customization in a machine shop are marked red.

Note: Excitation and detection objectives' optical axes are about 135 mm above the optical table.

Right excitation arm

The right excitation arm is identical to the left arm.

Switching illumination between arms

The laser is directed into one of the arm using internal switching mechanism - a flip mirror inside the laser combiner, which allows the latter to rapidly switch between its two fiber ports.

Detection arm

Optical design

Detection arms consists of a camera with telecentric lens attached, and filter wheel with emission filters in front of the lens.

Mechanical design

[UPDATE with 3D printed filter wheel] Detection overview The lens mount (red) consists of two 3D printed parts bolted together and attached to the rail carriage with M6 bolts. The filter wheel (red) is 3D printed and motorized with Robotis Dynamixel servo, with 5 filter positions.

Sample stages

Sample stages overview 1 Sample stages overview 2

The sample stages assembly consists of two vertical linear stages (Y axis, ASI Dual-LS100-FTP, travel range 100 mm) connected together with a custom bridge (red, ASI mesoSPIM-Bridge). The X- and Z-axis stages are mounted on the bridge, and the rotation stage is mounted on top Z-stage. The sample (green) in a sample cuvette (or holder) is attached to the rotating stage shaft via kinematic base (Thorlabs KB25/M). The immersion cuvette with index-matched oil is placed on adjustable prism table (Radiant Dyes RD-PDT-S).

Clone this wiki locally