Running stage initiation procedure

step 1 - remote all stage inserts / anything obstructing stage movements

It's important to run this every time the stage is turned off. This procedure is run within NIS-elements

The stage is about to move to it's extremes in order to test it is working correctly and ensure NIS-elements consistently refers to the correct absolution position values of the stage. NIS-elements behaves unexpectedly with this Marzhauser stage - if the stage is turned off after a microscope restart and stage initiation is not run the microscope refers to the current stage location as the origin which is arbitrary.

step 2 - selecte 'Objective clearance' in device tab

first click on this

step 3 - selecte 'Initiate stage' in device tab

next click on this