The Conductivity measurement type allows you the perform (four-wire) conductivity measurements. This page serves to detail what means what on the configuration panel.
Name | Required? | Description |
---|---|---|
Ground | Only for SPA use | SMU channel to use as ground/common - only relevant if using an SPA |
Source-Drain | Always required | SMU channel to use for sourcing source-drain current and measuring source-drain voltage |
Source-Gate | Only if gate voltage set | SMU channel to use for sourcing source-gate voltage and measuring source-gate/leakage current |
Four-Point Probe 1 | Optional | Voltmeter channel to use for measuring a four-point probe voltage |
Four-Point Probe 2 | Optional | Voltmeter channel to use for measuring a second four-point probe voltage |
Thermometer | Optional | Thermometer channel to use for measuring temperature during the measurement |
In general the measurement works by sweeping over values of source-drain current then measuring the voltages that develop across the device as a result. (i.e. source current, measure voltage). The voltage value that is eventually used for fitting and extracting a value of conductivity will depend on how the instruments are configured. This is defined below in the step-by-step logic.
Before explaining the logic step-by-step, we shall define the following abbreviations:
SDC: Source-Drain Channel
SGC: Source-Gate Channel
FPP1: Four-Point Probe Channel 1
FPP2: Four-Point Probe Channel 2
- Switch off all configured channels (SDC, SGC, FPP1, FPP2)
- Configure all channels to use the specified integration time for measurements
- If gate voltage is enabled, set SGC to source specified source-gate voltage
- Generate list of current values to sweep over
- Set SDC to source first current value in list
- Turn on all channels
- Select first value in current list to be
I
- Set SDC to source
I
amps of current - Wait for the specified delay time
- Measure voltages:
- SDC voltage
- FPP1 voltage (if configured)
- FPP2 voltage (if configured)
- Determine voltage to use for conductivity calculation
- If neither FPP1 nor FPP2 are configured, use SDC voltage
- If only one of FPP1 or FPP2 is configured, use its voltage
- If both FPP1 and FPP2 are configured, use the difference between their voltages
- If values are left, select next value in current list to be
I
and go to step 8 - Turn off all channels
- End measurement
Conductivity measurements have three "groups" of parameters:
- Basic
- Source-Drain
- Source-Gate
This is where we configure the basic parameters of the measurement. These are:
Name: The name to give the measurement.
Integration Time: The integration time, in seconds, to use when measuring.
Delay Time: The time, in seconds, to wait between setting the source-drain current and measuring the resultant voltages.
This is where we configure the set of current values to sweep over for the measurement.
Start Current: The current value to start the sweep at, in Amps.
Stop Current: The current value to end the sweep at, in Amps.
No. Steps: The total number of current values to use going from Start to Stop.
Scaling: How the current values in the sweep should be spaced. Linear means equal step sizes, for instance.
Sweep Both Ways: If checked, two current sweeps will be run. One going from Start to Stop followed by another going from Stop to Start.
This panel lets you optionally define a source-gate voltage to apply during the measurement.
Active: If checked, then the configured source-gate voltage will be used during the measurement.
Voltage: The source-gate voltage to use, in Volts - ignored if "Active" is not checked.