Reference is ISO 80601-2-61

Accuracy

  • Covered in section 201.12 of ISO 80601-2-61
  • How many beats before display
  • 2-4 beats (current market: 10-20s)
  • % accuracy:
    • 3% for SpO2 within 70-100%
    • 0.2% for Carboxy hemoglobin
    • Heart rate: 10 bpm

Display

  • Non-normalized waveform will indicate signal quality (201.12.4.102, see NOTE)
  • Waveform should run ~90% of the length of the display with ~25% of the height
  • Waveform should run 5 seconds of data (subject to change. May be 10 seconds)
  • Sample display: http://www.qmuniforms.com/photos/styles/DD006_1500_1.jpg

Physical

  • Dimensions of finger clip
  • Length of wire - 1-3 m
  • Dimensions of device - unconstrained
  • Weight - unconstrained

Electrical

Requirements

Compliance

Alarms