Reference is ISO 80601-2-61
- Covered in section 201.12 of ISO 80601-2-61
- How many beats before display
- 2-4 beats (current market: 10-20s)
- % accuracy:
- 3% for SpO2 within 70-100%
- 0.2% for Carboxy hemoglobin
- Heart rate: 10 bpm
- maximum: 30/sec
- minimum:
-
Temperature stability:
-
Shock and vibration (201.15.3.5.101.1): Amazing to have; would allow official use during transport
- Non-normalized waveform will indicate signal quality (201.12.4.102, see NOTE)
- Waveform should run ~90% of the length of the display with ~25% of the height
- Waveform should run 5 seconds of data (subject to change. May be 10 seconds)
- Sample display: http://www.qmuniforms.com/photos/styles/DD006_1500_1.jpg
- Dimensions of finger clip
- Length of wire - 1-3 m
- Dimensions of device - unconstrained
- Weight - unconstrained
- Must have: Battery-powered
- Must have: Ability to connect to continuous power
- Nice to have: Multiple abilities in one device
- UL, CE
- Nice to have: Button silences alarm x 5 minutes
- Remove reset button
- Add power switch
- Must have: Cut display if voltage makes readings unreliable
- Nice to have: Warning if voltage will soon become too low
- Allow test mode that tests each LED
- Can be activated by holding button for 20 seconds
- Report technical fault if 100% signal or 0% signal
See 201.13 (Hazardous situations and fault conditions)
Dashes or blinks to report technical fault (no finger; no reading; self-test failure)
Version 1.0: No alarms Version 2.0: SpO2 alarm (to be determined) Version 3.0: SPO2 alarm, CO alarm
R value will be taken and calculated (See FF.1)
See page 62 of ISO (Addendum FF.1)