# script to measure insertion loss with a power meter and signal generator
from pylab import *
from plx_gpib_ethernet import PrologixGPIBEthernet
import argparse
import pdb
from equipment.gigatronics_2520a import Gigatronics2520A
from equipment.agilent_e4418b import AgilentE4418B
if __name__ == '__main__':
parser = argparse.ArgumentParser(description='scalar network analyzer')
parser.add_argument('--cal', default = False, help='specify name of calibration file')
parser.add_argument('--ddir', default = './data/', help='specify name of calibration file')
parser.add_argument('filename', help='specify name of output file')
args = parser.parse_args()
prologix = PrologixGPIBEthernet('192.168.1.128', timeout = 10)
prologix.connect()
meter = AgilentE4418B(prologix = prologix)
synth = Gigatronics2520A(prologix = prologix)
freqs = np.arange(1e9, 20.1e9, .25e9)
synth.set_cw_power(0)
synth.output_on()
meas = np.empty_like(freqs)
for (i,f) in enumerate(freqs):
synth.set_cw_frequency(f)
meas[i] = meter.read_power(f)
print('f: {}, p: {}'.format(f/1e9, meas[i]))
cal = False
with open(args.ddir + args.filename, 'w') as datafile:
if args.cal:
with open(args.ddir + args.cal) as calfile:
cal = np.load(calfile)
assert np.array_equal(cal['frequencies'], freqs)
meas -= cal['power']
np.savez(datafile, frequencies=freqs, power=meas, calfile=args.cal)
plt.plot(freqs, meas)
plt.savefig('sweep.png')
synth.output_off()