Peltier status label is wrong #11

Open
maripo opened this issue Nov 10, 2018 · 0 comments
Open

Peltier status label is wrong #11

maripo opened this issue Nov 10, 2018 · 0 comments

Comments

@maripo
Copy link
Collaborator

maripo commented Nov 10, 2018

Label on console says "Peltier is heating" even if it's actually cooling.
It's caused because the firmware is sending "absolute" values of Peltier drive instead of signed values.